Ice-free anemometer: laboratory and field testing / S.G. Krishnasamy, J. Motycka.

Title: Ice-free anemometer: laboratory and field testing / S.G. Krishnasamy, J. Motycka.
Author(s): Krishnasamy, S. G.
Motycka, J.
Date: 1991.
Publisher: Ottawa: Environment Canada
In: Proceedings of the Third International Workshop on the Atmospheric Icing of Structures, 6-8 May 1986, Vancouver, Canada. (1991.),
Abstract: Instrument tested in wind tunnel. Tests indicate that 25 W heater is adequate to keep anemometer ice-free during freezing rain, but 50 W heater will be required at very low temperatures. Instrument being tested underestimated wind speeds when these were higher than 10 m s⁻¹.
Notes:

In: Proceedings of the Third International Workshop on the Atmospheric Icing of Structures, 6-8 May 1986, Vancouver, Canada / Leslie E. Welsh, D.J. Armstrong, eds.

Keywords: 551.5 -- Meteorology.
551.508.5 -- Wind, meteorological instruments.
624.143.54 -- De-icing methods, thermal.
D -- Atmospheric sciences.
SPRI record no.: 113051

MARCXML

LDR 01343naa#a2200000#a#4500
001 SPRI-113051
005 20240328125046.0
007 ta
008 240328s1991####onca###|#####|0||#0#eng#d
035 ## ‡aSPRI-113051
040 ## ‡aUkCU-P‡beng‡eaacr
100 1# ‡aKrishnasamy, S. G.
245 10 ‡aIce-free anemometer :‡blaboratory and field testing /‡cS.G. Krishnasamy, J. Motycka.
260 ## ‡aOttawa :‡bEnvironment Canada,‡c1991.
300 ## ‡ap. 311-318 :‡bill., diags.
500 ## ‡aIn: Proceedings of the Third International Workshop on the Atmospheric Icing of Structures, 6-8 May 1986, Vancouver, Canada / Leslie E. Welsh, D.J. Armstrong, eds.
520 3# ‡aInstrument tested in wind tunnel. Tests indicate that 25 W heater is adequate to keep anemometer ice-free during freezing rain, but 50 W heater will be required at very low temperatures. Instrument being tested underestimated wind speeds when these were higher than 10 m s⁻¹.
650 07 ‡a551.5 -- Meteorology.‡2udc
650 07 ‡a551.508.5 -- Wind, meteorological instruments.‡2udc
650 07 ‡a624.143.54 -- De-icing methods, thermal.‡2udc